admin管理员组文章数量:1537267
2024-07-29 作者:
Standard Operating Procedure of X-ray CT
(Nikon XTH225)
There are three main steps for performing X-ray CT on a sample. These steps are as follows:
I.
Inspect-X – CT Data Acquisition
The Inspect-X interface is usually opened on the main computer (Acquisition Computer) as shown
below.
In case if the interface was closed by the previous user, double-click on the Inspect-X icon on the
screen to start it. The startup window will look as shown below; click “Run” to continue:
Note: No password is required unless you are the equipment manager and would like to perform
maintenance or filament change.
After the startup, you will see the following window on the screen.
The system must be homed when it is first powered on. As the manipulator will move during this
procedure, always make sure that there is nothing on the table or in its way before homing the
system.
Then click “Homing” tab. After homing is completed, the red exclamation mark by
“Homing” icon should change to a green check mark.
For X-ray CT data acquisition follow the steps below:
1. Auto-condition
Verify that the system has been auto-conditioned before you start the scan. This step is necessary if you are the first user of the day.
To perform auto-condition, go to Setup and then click on “auto-condition” and wait until the
source is stable at the maximum beam energy (225 kV) for at least 5 minutes.
2. Mount the Sample
Securely mount the sample using hot glue. Tilt the sample, if possible, to minimize cone beam
artifacts in circular scan.
3. Choose a Scanning Method
Go to CT workflow as highlighted in the image below and click “Start” tab to choose a new scan.
For a small size sample, use “New Circular CT scan”. If you have a tall sample or a stack of
samples, a “New Helical CT Scan” is recommended. To repeat a scan using previous parameters,
you may choose “Use Parameters of Last Scan”.
4. Position the Sample and Set the Scanning Parameters
Use “Position and optimize” tab to position the sample and set the X-ray and detector parameters.
Follow the steps bellow:
• Use the joysticks on the control panel to zoom in and out as well as to rotate the sample.
• Always leave a thumb between the part and the edges to eliminate artifacts that can occur due
to missing voxel data.
• For optimal results, zoom in the sample to the max magnification possible, which results in
minimum effective pixel size.
Make sure the entire part remains within the image window throughout the scan.
• Rotate the sample so that the X-rays pass through the path of greatest attenuation while
setting up the parameters.
• Optimize X-ray parameters. To increase penetration, especially for dense materials like
Inconel and Stainless Steels, a very high beam energy is required. From the image histogram,
optimize a range of Gray Values (GV) so that the min GV is around 10k to ensure penetration
and the max GV is 50 – 60k (Background image/air). Make sure that the background is not saturated. You may need to use filters to satisfy this condition.
Note: The two factors defining the resolution of the image are the X-ray spot size and effective
pixel size. As a defocus threshold, make sure that the wattage is always ≤ 4 × (effective pixel size) for a rotating target and 1:1 for a static target.
To keep system power under defocus value, increase exposure and lower power (if possible)
(smaller spot size = sharper image).
Increasing the gain and exposure can help adjusting gray value alongside the power. However,
high exposure time increases the scanning time while high gain introduces some noise.
5. Reconstruction (skip this tap unless you would like to do automatic reconstruction, which is not
recommended as it does not always give the best results)
6. Acquire Dataset
In the “Acquire Dataset” tab, do the following steps:
• Name your sample: it is recommended that you give a meaningful name to your scan because
the name will appear on every projection, and it will make it easier for future reference.
• Create shading correction: in the shading correction field, increase the number of “Frames to
average” until the time is ~ 2:30 seconds. Leave the “Number of images” equals to 2, as the
flat panel detector responses linearly to X-ray energy. Leave the “White target” equals to
60k or set it equals to the average grey value from the histogram when the sample is not in
view.
• Minimize ring artifacts: selecting “Minimize ring artifacts” will improve scan quality, but it
will increase the scan time.
• Projections: increasing the “Number of projections” will improve quality of the scan;
however, there is a limit where there will be no more significant improvement when
exceeding the limit. An insufficient number of projections will result in aliasing effect in the data. Use the following image as a reference:
• Frames per projection: increasing the number of frames per projection will improve
resolution. However, increasing the number of frames also increases the scan time. A
practically suitable number of frames will be 2- 4 frames per projection. See the reference
figure below:
• Safe position: set a safe position so that the sample is moved out of the view for shading
correction. This can be done as shown below:
• Then, select the reconstruction computer, where the acquired data will be sent using CT agent
and click acquire to start collecting data.
Total Scan Time= (Exposure time) * (Projections) * (Frames)
II. CT Pro 3D
The data collected from Inspect-X will be stored in the “Active Project” drive on the reconstruction
computer. Go to CT data, locate your data folder, and double click on the CT Pro 3D icon, which will open the data using CT Pro 3D software. The interface window will look like the image below:
1. Image Tab
• Check Sample Movement
In the image tab, use the keyboard arrow keys to move back and forth between the first and the
extra image to make sure that the sample has not moved between images as well as there is no
change in brightness between images.
• Fix Variation in Image Brightness (rarely required)
Within the image tab, variation in brightness can be fixed by applying flux normalization to a
small region of the image, where the part does not rotate through. To perform this action, follow the steps shown in the image below.
2. Center of Rotation Tab
The center of rotation is calculated using eighter a single or dual slice setting. This step can also be done in the automatic mode, through which the software finds the pixel shift based on the
selected row(s) of pixels from the tiff images. It is recommended to use a dual slice setting for tall
samples. Using a single slice will not correct for angularity in the panel alignment.
Follow the steps below to calculate the center of rotation:
3. Setup
The “Setup” tab includes many features as shown in the image below. However, unused features
are usually minimized, and the most used features are shown in the following images.
Below are the most used features (details follow on the next page)
• Minimize Volume Size (not necessary if you have a big storage)
• Fix Beam Hardening
For beam hardening, select “reconstruct selected” to perform all beam hardening preset (apply different attenuation curves).
• Reduce Noise
Not always needed, but it can be easily done as follows:
4. Calibration Tab
Calibration is only used in medical field (This tab is deactivated on our CT Pro 3D)
5. Volume Tab
The default settings are the optimized settings for most scans. The only setting that needs to be
adjusted is the volume box to reduce the size or choose the part of the scan that you need to process.
III. VGStudio
After the reconstruction is done and the data are converted, a new folder will be added to the data. The new folder is VGStudio compatible and VGStudio icon will appear in the folder as shown below:
1. Double click on the VGStudio icon and you should see the following window:
2. In the rendering window, you will find two lines as shown in the image. Move the red vertical line
“isosurface line” to the mid-distance between the two humps to separate material from the
background and the blue diagonal line to adjust opacity. You may add a couple more joints to the
opacity line by clicking “+” sign at the bottom for convenience.
In the image bellow both lines are adjusted, and the background of the slices is changed to Black for
better visualization.
3. Perform surface determination. Right click on the volume and choose surface determination as shown
below:
In the surface determination window, set the Approach to “advanced classic” and material determination
to “manual” or “automatic”, depending on your analysis. Then, in the options at the bottom of the
window, set starting contour healing to “Remove all particles and voids” and click finish. The new surface determination window with all the above settings will look like the image below:
4. Fit three geometries to your object to use as references for its registration. This step can be skipped if
you have a CAD model to use for registration or if you decided to use simple registration.
5. Registration
Registration is required for analysis, so it is necessary to register your object. Use one of the registration
methods from the list by clicking on the registration icon shown in the image. For example: “3-2-1
registration” is usually used when you have fitted three geometries to the object. “Best fit registration” is used when you have a CAD model.
Below is an example of simple registration window. Grab the middle red square shown in the window
below to move the object up and down while click the object and hold to rotate the object. When you have aligned your object, click finish to complete the registration.
6. Analysis
Now you are ready to perform analysis. As an example, porosity analysis is explained blow.
• Click the porosity analysis icon shown below and choose the type of porosity analysis. The
second in the list is used for this example.
• In the next window, choose your algorithm (VGDefx or only threshold). “Only threshold” is
a very fast algorithm in comparison with the other algorithms. However, some algorithms do better than the others depending on your scan.
• For defining the “Void max.”, highlighted in yellow, use the eye dropper to select the
background and the actual material. First, select a small area of the background in one of the
slice windows, which will appear as a line in the other windows. Drag the line to extend the
selected area to a small volume and click “Next”. Do the same for defining the material. It is necessary to avoid pores as you select your material.
• Adjust the rest of the settings as required and click “Calculate”.
• The analysis window of this example is shown below.
Zoom any area in or out by scrolling the middle roller of your mouse while holding down the control key on the keyboard.
Scroll the middle-roller of the mouse to move through the slices in any of the windows.
• Change the transparency in the histogram window to view the volumetric porosity content in the
3D object as shown.
Please note that VGStudio is a very powerful software, which requires additional training for advanced
analysis. What has been presented here is only to get you started and do only basic analysis.
本文标签: 尼康XTH225CT扫描仪标准操作规程说明书
版权声明:本文标题:尼康XTH225-CT扫描仪标准操作规程说明书 内容由热心网友自发贡献,该文观点仅代表作者本人, 转载请联系作者并注明出处:https://m.elefans.com/shuma/1722185358a919365.html, 本站仅提供信息存储空间服务,不拥有所有权,不承担相关法律责任。如发现本站有涉嫌抄袭侵权/违法违规的内容,一经查实,本站将立刻删除。
发表评论